5TypesofMeasurements
Settings for Patient ID Input Interface 
In the patient ID input interface menu, you can perform settings related to the position of the start 
byte for reading data, number of bytes for reading input data, method for inputting data, and 
connection tests for a USB device for inputting the patient ID.
How to Enter the Patient ID Input Interface Menu 
MENU screen > 
 Input tab > [Input] 
Setting items 
Description 
Factory 
settings 
Update
Set the input methods to handle a USB device for automatic input 
(magnetic card reader, barcode reader, etc.) and a USB device for 
manual input (numeric keypad, etc.). 
[OVERWRITE]: Input data overwrites existing data. Select when a USB 
device for automatic input is connected. 
[ADD]: Input data is added to existing data. Select when a USB device 
for manual input is connected. 
ADD
Start Position
Set the position of the byte at which to start reading data that is input 
from a USB device. 
You can only do this setting if [OVERWRITE] is selected for [Update]. 
[0 to 255]: Position of byte at which to start reading 
0
Data Length
Set the number of bytes to read of the data that is input from a USB 
device. The length of data (number of bytes) read from the reading start 
byte position set for [Start Position]. 
You can only do this setting if [OVERWRITE] is selected for [Update]. 
[1 to 64]: Length of data to read 
1
Connection 
Check
For this setting, press the measurement button to do a connection test 
for the USB device and display the results. 
If the connection is normal, a message to check at USB connection 
check screen appears. If the connection is not normal, an error message 
appears. 
[YES]: Go to the USB connection check screen. 
[NO]: End the USB connection test. 
Settings for Interface of Output of Measurements 
You can send measurement results to an external device via LAN or RS-232C. 
 CAUTION
•  Always turn off the power before connecting cables. 
•  Use the computer and other equipment that conform to the system standard IEC 60601-1 or 
IEC 60950-1 for the Full Auto Tonometer TX-20. Be sure that the entire system conforms to 
IEC 60601-1-1. Be sure to also use an isolation transformer conforming to IEC 60601-1 when a 
computer conforming to IEC 60950-1 is used.  
Otherwise, electric shock may result. Ask your sales representative or local Canon dealer to 
connect it to a computer.
•  Be sure to attach the provided ferrite core before connecting the LAN cable to the LAN 
connector. 