7Troubleshooting
Message and Remedy
The message screen appears if a problem occurs or if there is a notification. 
If this happens, check the displayed message, refer to the message codes and error codes on the 
following list and follow the instructions in the remedy. If the message screen does not close or if a 
message not shown in the table is displayed, contact your sales representative or local Canon dealer.
Message content
Displays the content of the notification and the 
status of the problem that has occurred. It may 
also display the error code which indicates the 
details of the error status.
Message code
The message category is indicated by the first 
letter.
Fxxx: The customer cannot resolve this error.
Exxx: The customer may be able to apply a remedy 
to resolve this error.
Wxxx: Alerts the customer to be careful.
Check icon
Message Screen Operations 
1
Check the message. 
2
Select the check icon. 
Rotate the measurement unit vertical movement ring to select the confirm icon, and press the 
measurement button.
3
Apply the remedy. 
Apply the remedies as shown in the message. 
1
2
3