7Troubleshooting
Warning
Code
Cause 
Remedy 
320
Because of a dead battery (for backup), automatic 
updating of date stops.
1)  Measurements can be done. Select either one from the 
followings and use the TX-20 as it is.
•  Select [OK]. 
Reflect the clock on the measurement results. Set the 
correct clock again on the displayed clock input screen. 
However, the clock you have set again is valid only when 
the power is turned on.
•  Select [CANCEL]. 
The clock is not reflected correctly on the measurement 
results. The clock input screen is not displayed and the 
TX-20 is reset at the initial value 2010/1/1.
If this error occurs, the clock function of the main unit is not 
kept. Therefore, you need to set the clock each time you 
turn on the power. We recommend replacing the battery. For 
replacing the battery, ask
 your sales representative or local 
Canon dealer.
329
The USB device of full-speed specifications was connected 
to this product.
1)  Record the message code, contents (including error 
codes) and then execute [OK]. 
2)  Connection check is necessary. Ask your sales 
representative or local Canon dealer to perform the 
connection check. 