Guidance and Manufacturer’s Declaration for Electromagnetic Immunity
The TX-20 is intended for use in the electromagnetic environment specified below.
The customer or the user of the TX-20 should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic Environment – Guidance
Electrostatic discharge 
(ESD)
EN IEC 61000-4-2
±(2, 4, 6) kV contact
±(2, 4, 8) kV air
±(2, 4, 6) kV contact
±(2, 4, 8) kV air
Floors should be wood, concrete or ceramic 
tile.
If floors are covered with synthetic material, 
the relative humidity should be at least 30%.
Electrical fast transient/
burst
EN IEC 61000-4-4
±2 kV for power supply 
lines
±1 kV for input/output 
lines
±2 kV for power supply 
lines
±1 kV for input/output 
lines
Mains power quality should be that of a 
typical commercial or hospital environment.
Surge
EN IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality should be that of a 
typical commercial or hospital environment.
Voltage dips, short 
interruptions and 
voltage variations on 
power supply input 
lines
EN IEC 61000-4-11
<5% U
T
 (>95% dip in 
U
T
) for 0.5 cycles
40% U
T
 (60% dip in U
T
for 5 cycles
70% U
T
 (30% dip in U
T
for 25 cycles
<5% U
T
 (>95% dip in 
U
T
) for 5 sec
<5% U
T
 (>95% dip in 
U
T
) for 0.5 cycles
40% U
T
 (60% dip in U
T
for 5 cycles
70% U
T
 (30% dip in U
T
for 25 cycles
<5% U
T
 (>95% dip in 
U
T
) for 5 sec
Mains power quality should be that of a 
typical commercial or hospital environment. 
If the user of the TX-20 requires continued 
operation during power mains interruptions, it 
is recommended that the TX-20 be powered 
from an uninterruptible power supply.
Power frequency 
(50/60Hz) magnetic 
field
EN IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be 
at levels characteristic of a typical location in 
a typical commercial or hospital environment.
NOTE:  U
T
 is the AC mains voltage prior to application of the test level.