FDR nano has a low output in order to downsize, but it can be captured
without any problems by D-EVOII and Virtual Grid.
Comparison with FDR Go
General statement
Item
Maximum mAs
Tube current
Minimum exposure time
Focus size
FDR nano spec
(D-EVO II+VG)
25 mAs
25 to 35 mAs
10 ms
1.2 mm
Image concern
(1)Sortage of dose
(2)Mortion artifact
(3)Neonate radiography
(4) Enlarged blur
(Reference) FDR Go spec
(D-EVO + mobile X-ray system)
320 mAs
100 to 400 mAs
1 ms
0.7/1.3 mm
System
FDR nano
FDR Go
DR
D-EVO II
VG
D-EVO
X-ray equipment
X-ray Cart(2.5 kW)
Mobile X-ray system(32 kW)
FDR Go
FDR nano
For image quality at the output of