Guidance and manufacturer’s declaration - electromagnetic immunity
The UP-DR80MD is intended for use in the electromagnetic environment specified below. The customer or the user of the 
UP-DR80MD should assure that it is used in such as environment.
Immunity test
IEC 60601 test 
level
Compliance level
Electromagnetic environment-guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 Vrms
3 V/m
Portable and mobile RF communications equipment should 
be used no closer to any part of the UP-DR80MD, including 
cables, than the recommended separation distance 
calculated from the equation appliance to the frequency of 
the transmitter.
Recommended separation distance
d = 1.2 
P
d = 1.2 
P   80 MHz to 800 MHz
d = 2.3 
P   800 MHz to 2.5 GHz
Where P is the maximum output power rating of the 
transmitter in watts (W) according to the transmitter 
manufacturer and d is the recommended separation distance 
in meters (m).
Field strengths from fixed RF transmitters, as determined by 
an electromagnetic site survey, a should be less than the 
compliance level in each frequency range. b 
Interference may occur in the vicinity of equipment marked 
with following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 
reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile 
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To 
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be 
considered. If the measured field strength in the location in which the UP-DR80MD is used exceeds the applicable RF 
compliance level above, the UP-DR80MD should be observed to verify normal operation. If abnormal performance is 
observed, additional measures may be necessary, such as reorienting or relocating the UP-DR80MD.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.