Emerson 7003M 用户手册

下载
页码 74
Operator Manual
748223-K
June 2002
3-12     Operation
Rosemount Analytical Inc.    A Division of Emerson Process Management
Model 7003M
ALARM1
ALARM2
%O2
°C
SET-UP Mode
+
SECURITY ACCESS ROUTINE
(see Figure 3-4)
ALARM1
ALARM2
%O2
°C
ALARM1
ALARM2
%O2
°C
ALARM1
ALARM2
%O2
°C
Flashing
May not be displayed
each time.
ALARM1
ALARM2
%O2
°C
Flashing
Prompts for Security Code.
If left at 000, code will be disabled.
Use Arrow keys to set Security Code.
ALARM1
ALARM2
%O2
°C
Select either 4-20mA or
0-20mA Output
ACK
Prompts for low end output.
(i.e., 0% O
2
 would be the low range
representing O
2
 level of 0 or 4mA.
ENTER
CAL
ENTER
ENTER
ALARM1
ALARM2
%O2
°C
Flashing
ALARM1
ALARM2
%O2
°C
ENTER
Flash for 2 seconds
(Hi range output selection)
ENTER
ALARM1
ALARM2
%O2
°C
Set Span output level = 20mA
Flashing
May not be displayed
each time.
Flashing
ALARM1
ALARM2
%O2
°C
ENTER
Set Span output level = 20mA
ALARM1
ALARM2
%O2
°C
Flash for 2 seconds
Prompts for zero offset
ALARM1
ALARM2
%O2
°C
ALARM1
ALARM2
%O2
°C
Figure 3-3. SETUP Mode Flowchart