Emerson NGA2000 用户手册

下载
页码 178
6 - 28
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Main Menu — Technical level configuration
↓↓
Diagnostic menus
↓↓
Analyzer module diagnostics
↓↓
Power supply voltages
↓↓
In the submenu "Power supply voltages" you can find the current value of the power
supplies ("is"-lines) and the values when the CLD analyzer module was manufactured
("was"-lines).
HOME
ESCAPE
MORE
INFO
CLD
F1
F2
F3
F4
F5
95.0 ppm
Analyzer diagnostics
+15V analog is:
15.21 V
-15V analog is:
-15.21 V
-15V analog was:
-15.11 V
+5V digital is:
5.010 V
+5V digital was:
5.007 V
23.82 V
+12V analog is:
11.97 V
+12V analog was:
12.05 V
+15V analog was:
15.02 V
+24V-power was:
24.22 V
Power supply voltages
6.3.2.1 Diagnostic Menus
Power Supply Voltages