Main Menu — Expert controls and set up
↓↓
Analyzer module set up
↓↓
Calibration parameters
↓↓
In the menu "Calibration Parameters" you can set up several parameters of the zero and
span calibration.
Set up parameters:
♦ 
Select the line of variables you want with the 
↓↓
 -key or the 
↑↑
 -key.
♦ 
Select the variable with the ENTER key or the 
 -key.
♦ 
Select any digit with the 
 -key or the 
 -key and adjust a new value with
the 
↑↑
 -key or the 
↓↓
 -key
resp. select the whole parameter with the 
↑↑
 -key or the 
↓↓
 -key.
♦ 
Confirm the new value with the ENTER key or
cancel and go back to the last value with the F2 key.
Line of variables "Calibration adjustment limits:
♦ 
Enabled: The calibration result will only be accepted within the limit settings. If the
results will be out of these limits, you will find a corresponding error message in the
menu "Analyzer zero" respectively "Analyzer span". Illustrations of these menus you will
find in chapter 4.2.1 p. 4-22 resp. 4.2.2 p. 4-26.
♦ 
Disabled: The calibration will be possible, even if the calibration results will leave the
allowed limits. Compare with the explanations in chapter 5.1.1 p. 5-6/7.
HOME
ESCAPE
INFO
CLD
F1
F2
F3
F4
F5
95.0 ppm
Calibration Parameters
Calibration adjustment limits:
Disabled
Calibration failure alarm:
No
Cal. failure error allowed:
50 %
Calibration time out:
60 s
Zero ranges:
TOGETHER
Calibration averaging time:
5 s
Span ranges:
SEPARATELY
5.1.2 Analyzer Module Set up
Calibration Parameters