© Adam Equipment Company 2011 
Page  
32
 
 
 
 
10.0 CALIBRATION 
 
ACTION 
DISPLAYS 
Press the [Tare/Zero] key during the self-
test at power on. The scale will ask to 
enter the password. 
 
 
“ Pi n” 
Default Password is 0000. 
Enter “0” four times. 
Password can be changed in technical 
parameters. 
Press [Print]
 
 
“ Pi n”     “ - - - -” 
Select the scale to be set up by using the 
[Local/Remote] key. 
Press the [Print] key to enter the 
technical section.  
 
“tECH”  “   LocAL”   “      “   
“tECH”  “rEmo tE”   “      “   
 
Use the [U.Wt.] to select the weighing unit 
to be used for setting  the calibration for 
the scale. The arrow in the “Weight” 
window will indicate the unit selected. 
Press the [Print] key to continue. 
 
  
“tECH”  “   Uni t”   “      “   
 
 
 
You will enter the first parameter -
Calibration. Display will ask to unload any 
weight on the platform. 
Press the [Print] key to continue. 
“nO L o Ad” 
 
 
 
   
Load the calibration weight onto the scale  
 
If the mass is different than the value 
shown press the [CE] key then enter the 
calibration weight to be used . 
 
Then press [Print]
 
“LoAd   ”    “ 0010”    “ KiLoS”