SECTION 5—OPERATING INSTRUCTIONS
22
2. If the flowrate on the flowmeter ever falls below 0.5 L/min 
for Perfecto
2
 5 for more than about one minute, the LOW 
FLOW alarm will be triggered. This is a rapid beeping of 
the audible alarm. Check your tubing or accessories for 
blocked or kinked tubing or a defective humidifier bottle. 
After rated flow is restored, more than 0.5 L/min for 
Perfecto
2
 5 , the LOW FLOW audible alarm will go off.
NOTE: The use of some accessories such as the pediatric flowstand 
and the HomeFill compressor will deactivate the Low Flow Alarm.
FIGURE 5.5    Flowrate
SensO
2
 Oxygen Purity Indicator (If Your 
Unit Features the O
2
 Sensor Feature)
This feature monitors the purity of oxygen generated by the 
oxygen concentrator. If purity falls below factory preset 
standards, indicator lights on the control panel will 
illuminate.
Flow Knob
Ball
Flowmeter
Front Panel
0.5 L/min