Invacare IRC5PO2 用户手册

下载
页码 32
SECTION 6—MAINTENANCE
25
SECTION 6—MAINTENANCE
NOTE: The Invacare concentrators are specifically designed to 
minimize routine preventive maintenance. In places with high dust 
or soot levels, maintenance may need to be performed more often. 
Only qualified personnel should perform preventive maintenance on 
the concentrator. Power should be disconnected before beginning 
preventive maintenance on the Invacare concentrator.
NOTE: At a minimum, preventive maintenance MUST be 
performed according to the maintenance record guidelines. Refer to 
Preventive Maintenance Record on page 27.
Routine Maintenance
NOTE: The following routine maintenance should be performed 
more frequently. See recommended intervals in each section and 
preventive maintenance record. 
Cleaning the Cabinet Filter
NOTE: For this procedure, refer to FIGURE 6.1 on page 26. 
NOTE: There is one cabinet filter located on the back of the cabinet.
1. Remove the filter and clean at least once a week depending 
on environmental conditions.
NOTE: Environmental conditions that may require more frequent 
cleaning of the filters include but are not limited to: high dust, air 
pollutants, etc.
2. Clean the cabinet filter with a vacuum cleaner or wash in 
warm soapy water and rinse thoroughly.
3. Dry the filter thoroughly before reinstallation.
CAUTION
DO NOT operate the concentrator without the filter 
installed.