mitutoyo 501s 用户手册

下载
页码 48
 
 
 
DESIGNING THE FIXTURES 
 
This chapter describes precautions to be observed when attaching 
the emission unit and the reception unit, which have been detached 
from the support base of the measuring unit, to a specially arranged 
dedicated fixture. 
 
 
 
4.1 Outline 
 
In application, the emission unit and the reception unit may have to be detached from the 
support base of the measuring unit and attached to a dedicated fixture. If this is the case, 
the measuring accuracy cannot be ensured unless the emission unit and the reception unit 
are properly aligned on the dedicated fixture. Design a proper fixture according to this 
section. 
 
4.2  Consideration to Calibration 
 
To ensure accurate measurements, design the fixture so that the workpiece to be 
measured can be positioned at the point of measurement that is the focal point of the 
emission unit. Also, make allowances for calibration and for a calibration gage to be 
mounted on the fixture. 
 
4.2.1 Measuring 
position 
and resulting accuracy 
As shown in the figure below, the scanning beam of the measuring unit is produced by 
reducing the thick beam to a beam of the minimum diameter at the measuring position that 
is the focal point. 
However, since the inclination of the reception signal is defined as “beam 
diameter/scanning speed”, the inclination will be the steepest at measuring point (b) and will 
be less steep at points (a) and (c), off from the measuring point. 
The less steep the inclination of the reception signal, the more susceptible the signal is to 
noise and light disturbance, resulting in degraded repeatability. Thus, due attention should 
be paid to ensure that the workpiece is located at the measuring position. 
 
 
 
 
 
 
 
 
 
 
 
Measuring position that is focal point. 
Inclination of the reception 
Measuring region
Emission unit side
(a)          (b)      (c)
Reception unit side 
No. 99MBC094A
 
4-1