GE PT878GC 用户手册

下载
页码 308
Programming Site Data
4-51
April  2004
Entering Reynolds Correction  (cont.)
b. Press 
[ENTER]
 to confirm your selection.
Note: If you disable Reynolds Correction, you will not be able to 
enter any other values in this window.
2. If you enable Reynolds Correction, you will also need to enter 
the Kinematic Viscosity of the gas (available in the brochure 
Sound Speeds and Pipe Size Data, 914-004). The next prompt 
asks if you wish to enter the Kinematic Viscosity as a single 
number, or as a table of values.
a. Use the  [
W
 ] and [
X
] keys to scroll to the appropriate 
radio button.
b. Press 
[ENTER]
 to confirm your selection.
3. The following steps depend on whether you select a single 
value or a table for Kinematic Viscosity. 
If you entered Single:
The prompt asks for a kinematic viscosity value.
a. Press 
[ENTER]
 to open the text box.
b. Use the numeric keys to enter the known value (from 
Sound Speeds and Pipe Size Data).
c. Press 
[ENTER]
 to confirm your selection.
You have completed entering data for a single Kinematic 
Viscosity factor.
If you entered Table:
The menu asks, first for the data source, and then for the entry of 
values into the Reynolds Correction table. This procedure allows 
the PT878GC to estimate Kinematic Viscosity from other 
measurements, such as signal strength.
a. Press 
[ENTER]
 to open the Data Source window. The 
Select Measurement window (Figure 4-19 on page 4-40) 
appears.