Programming Meter Settings
7-5
April  2004
The Battery Charger
The Battery option allows you to monitor the current run time and 
status of the internal rechargeable batteries, as well as to 
condition NiCad batteries to maintain the maximum life possible. 
Conditioning NiCad batteries (a process that can take up to 12 
hours for a fully charged pack) discharges the pack completely 
and then performs a fast charge. 
Note: When conditioning the batteries, be sure you have 
plugged the AC adapter into the PT878GC and pressed 
the power key. NiMH batteries normally do not require 
conditioning.
To open the option window:
1. From the Meter menu, scroll to the Battery entry and press 
[ENTER]
. The Battery Charger window opens, as shown in 
Figure 7-3 below. 
2. If you wish to open the Part Number window, press 
[ENTER]
You can choose from a 1.8 Ahr NiCd battery (part number 
200-058) or a 3.0 Ahr NiMH battery (part number 200-081). 
Scroll to the battery type you have installed and press 
[ENTER]
.
Note: If you do not set the correct battery type in the Part 
Number window, the battery continues to function, but the 
battery status icons (see page 2-5) will not be accurate.
Figure 7-3:  The Battery Charger Window
Battery Charger
Cancel
l
OK
Status: Full Charge
Battery Voltage: 5.78 Volts
Battery Type:  NiCd
Condition Battery
Run Time Remaining:  
Backup Battery: OK
       Part Number:
1.8Ahr NiCd (200-058)