Servicing the PT878GC
Programming the Thickness Gauge (cont.)
1. Press the [
T
] arrow key to enter the window.
2. The first prompt asks for the low signal threshold. If the signal 
strength falls below this value, a “Low Signal” message 
appears on the screen. To change this value:
a. Press 
[ENTER]
 to open the text box.
b. Use the numeric keys to enter the provided value.
c. Press 
[ENTER]
 to confirm your entry.
3. The next value, transducer delay, represents the time the 
ultrasonic signal takes to travel from the meter to the pipe 
material surface. To change this value:
a. Press 
[ENTER]
 to open the text box.
b. Use the numeric keys to enter the desired value.
c. Press 
[ENTER]
 to confirm your entry.
4. The signal inversion prompt enables you to invert the 
thickness transducer signal. Signal inversion may be 
necessary if your thickness readings show unusual jitter, or for 
certain unusual combinations of pipe and fluid or pipe and 
lining. Consult the factory before measuring the thickness of 
lined pipes.
Note: When measuring the thickness of certain pipe materials, 
you might need to drain the pipe before measuring the 
pipe wall, because certain pipe/liquid combinations 
muffle the ultrasonic echo.
Signal inversion should be on, unless you have received other 
instructions. To change its status, use the [
W
 ] and [
X
] arrow 
keys to move to the appropriate radio button and press 
[ENTER]
.
5. To enter the noise threshold: