OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
POWER CONNECTION
HI 983315-0 and HI 98331500
HI 983315-0 and HI 98331500
HI 983315-0 and HI 98331500
HI 983315-0 and HI 98331500
HI 983315-0 and HI 98331500
Connect a 2-wire power cable to the terminal strip while
paying attention to the correct positive and negative polari-
ties (12 VDC).
HI 983315-1
HI 983315-1
HI 983315-1
HI 983315-1
HI 983315-1
Connect a 3-wire power cable to the terminal strip while
paying attention to the correct earth, neutral and line
contacts (115/230 VAC).
ALARM CONTACT
Use this contact (maximum 5A, 240VAC, 30VDC) for connec-
tion to an alarm or dosing system. The unit acts as a switch
to control an external device.
Notes:
Notes:
Notes:
Notes:
Notes:
• All external cables connected to the rear panel should
   end with wire lugs.
• It is recommended to cover the unused terminals with
   insulating tape.
• At start up the meter needs a few seconds to stabilize.
   Wait until a stable reading is displayed.
ACCESSORIES
ACCESSORIES
ACCESSORIES
ACCESSORIES
ACCESSORIES
HI 7634-00
EC/TDS probe
HI 70033P
84 µS/cm Calibration solution, 20mL
HI 7033M
84 µS/cm Calibration solution, 230mL
HI 7033L
84 µS/cm Calibration solution, 460mL
HI 7061M
Electrode cleaning solution, 230 mL
HI 710005
12 VDC power adapter, US plug
HI 710006
12 VDC power adapter, European plug
HI 710012
12 VDC power adapter, Australian plug
HI 710013
12 VDC power adapter, South African plug
HI 710014
12 VDC power adapter, UK plug
HI    731326
Calibration Screwdriver (20 pcs)
CALIBRATION
Make sure the meter is in the measurement mode (the
“MEAS” LED lights on).
Immerse the probe in HI 7033 (84 µS/cm) calibration
solution.
Shake briefly and wait for reading to stabilize.
Using a small screwdriver adjust the calibration trimmer
until the meter displays “42” ppm.
SETPOINT
Press the “SET” key. The display will show the default or
previously adjusted value for the setpoint.
Using a small screwdriver, adjust the “SET” trimmer until
the required limit value is displayed.
PROBE MAINTENANCE
To improve probe performance and prolong its life, it is
recommended to clean it regularly.
• Immerse the tip of the probe in HI 7061 Cleaning
Solution at least for one hour.
• If a more thorough cleaning is required, brush the
metal pins with very fine sandpaper.
• After cleaning, rinse the probe with tap water and
recalibrate the meter.
•  When not in use, clean the probe before storing it.
Front view of the
panel-mounted unit
Side view of the
panel-mounted unit
Adjustable location brack-
ets (supplied with the
meter) allow the controller
to slide into the cutout and
will hold the unit  securely
in place. 95 mm (3.74") is
the minimum amount of
space required to install
the controller.
Panel-mounted
unit assembling
view
73mm
2.87"
79mm
3.11"
42mm
1.65"
49mm
1.93"
95mm MIN
3.74"
76mm
3"
0.25/4mm
0.01/0.160"
79mm
3.11"
ADJUSTABLE
LOCATION
BRACKET
HI983315 LAYOUT
HI983315 LAYOUT
HI983315 LAYOUT
HI983315 LAYOUT
HI983315 LAYOUT
HI 740146
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FRONT PANEL
FRONT PANEL
FRONT PANEL
FRONT PANEL
FRONT PANEL
Keypad
SET
SET
SET
SET
SET
To display setpoint
MEAS
MEAS
MEAS
MEAS
MEAS
To display  measure-
ment
Trimmers
CAL
CAL
CAL
CAL
CAL
For calibration
SET
SET
SET
SET
SET
To adjust setpoint
LEDs
SET
SET
SET
SET
SET
ON when the LCD
displays  set value
MEAS
MEAS
MEAS
MEAS
MEAS
ON when the LCD displays  measured  value
ALARM
ALARM
ALARM
ALARM
ALARM
ON when the alarm contact is closed
REAR PANEL
REAR PANEL
REAR PANEL
REAR PANEL
REAR PANEL
1. Power Supply:
     HI 983315-0
     HI 983315-0
     HI 983315-0
     HI 983315-0
     HI 983315-0          HI 983315-1
          HI 983315-1
          HI 983315-1
          HI 983315-1
          HI 983315-1
     HI 98331500
     HI 98331500
     HI 98331500
     HI 98331500
     HI 98331500
             12 VDC             115/230 VAC
             12 VDC             115/230 VAC
             12 VDC             115/230 VAC
             12 VDC             115/230 VAC
             12 VDC             115/230 VAC
L1:
L1:
L1:
L1:
L1:
Positive                    230 VAC
L2:
L2:
L2:
L2:
L2:
Negative                  115 VAC
L3:
L3:
L3:
L3:
L3:
   ---                      Neutral
L4:
L4:
L4:
L4:
L4:
   ---                   Protection Earth
Note
Note
Note
Note
Note: : : : : the power input is internally protected by a 400 mA fuse
2. Alarm Contact: A1, A2.
3. Probe Connection: follow the above diagram and connect the
     colored wires of probe cable as indicated. It is recommended
    to connect the shield (P5) to avoid any interference.
OPERATING THE METER
All operations can be controlled via front panel keys and
trimmers.
“SET” and “MEAS” LEDs light up to indicate which is the
operating function.
Make sure that the meter is calibrated and the Setpoint is
properly selected  before performing any measurement.
Attach the probe to the meter. Install the probe in the
fittings or immerse it in the solution to be monitored, while
making sure that metal pins are completely submerged.
Press the “MEAS” key.
The LCD will show the TDS value of the solution in ppm unit.
Any initial variation on readings may be due to tempera-
ture compensation.
The “ALARM” LED will light up when the alarm contact is
closed, to indicate a TDS value higher than selected setpoint.