LP 2000 displays several different LCD codes to help the user to
operate the meter more efficiently.
This indicates that the meter is in a ready
state and measurement or calibration can be
performed.
This indicates that the Calibration Mode is
active. If the CAL key is not pressed within 6
seconds, the meter will automatically revert
back to the Measurement Mode.
This indicates that the Calibration Date Set-
ting Mode is active.
This indicates that a 0 FTU standard cali-
bration solution is required.
This indicates that the instrument is being
calibrated at zero.
This indicates that a 10 FTU standard calibra-
tion solution is required.
This shows the last calibration date (MM.DD).
This indicates that the sample is being ana-
lyzed.
This indicates that an error occurred during
calibration (for example a wrong standard
was used). Repeat the calibration with fresh
standards.
LCD AND ERROR CODES