Chapter 2
Function Generator Operation
©
 National Instruments Corporation
2-5
Figure 2-4.  Waveform Generation Trigger Sources
If you need to automatically trigger the waveform generation, use software 
to generate the triggers. A rising TTL edge is required for external 
triggering. For more information on triggering over RTSI lines, see the 
 section later in this chapter.
Modes of Operation
The NI 5401 has three triggering modes—single, continuous, and 
stepped—described in the following sections.
Single Trigger Mode
The waveform you define in the staging list is generated only once by going 
through the entire staging list. Only one trigger is required to start the 
waveform generation.
In single trigger mode, after the NI 5401 receives a trigger, the waveform 
generation starts at the first stage and continues through the last stage. The 
last stage is generated repeatedly until you stop the waveform generation. 
Figure 2-5 illustrates a single trigger mode of operation.
Software Trigger
RTSI/PXI Trigger
RTSI/PXI Trigger
Lines <0..6>
RTSI Switch
Trigger Select
Start Trigger
External Trigger
7
Digital
MUX