National Instruments SCC-TC01 用户手册

下载
页码 20
© National Instruments Corporation
7
SCC-TC Series Thermocouple Input Modules User Guide
You can use the following formulas to convert the cold-junction sensor 
voltage to cold-junction temperature:
where T
K
 is the temperature in kelvin.
where
a = 1.295361 
× 10
–3
b = 2.343159 
× 10
–4
c = 1.018703 
× 10
–7
R
T
 is the resistance of the thermistor in ohms.
where V
TEMPOUT
 is the output voltage of the cold-junction sensor.
where T(°F) and T(°C) are the temperature readings in degrees Fahrenheit 
and Celsius, respectively.
Note
V
TEMPOUT
 varies from 1.91 V (at 0 °C) to 0.58 V (at 55 °C). For best resolution, use 
the maximum gain for this range on the analog input channel.
For the cold-junction sensor measurement accuracy, refer to the 
Uncompensated Connectors and Accuracy
If you are using an SCC-TC02 with an uncompensated SCC panelette, 
temperature gradients between the module and the junctions on the 
panelette affect the accuracy of your measurements.
T
°C
( )
T
K
273.15
=
T
K
1
a b
R
ln
T
(
c
R
ln
T
(
)
3
+
+
[
]
--------------------------------------------------------------
=
R
T
5 000
V
TEMPOUT
2.5 V
TEMPOUT
--------------------------------------
,
=
T
°F
( )
T
°C
( ) ]9
[
5
------------------------