National Instruments NI 5102 用户手册

下载
页码 87
Chapter 4
Hardware Overview
© National Instruments Corporation
4-19
Figure 4-12. RTSI Bus Trigger Lines
The RTSI bus clock line is a special clock line on the RTSI bus that can only
carry the timebase of the master to the slaves. For the smallest jitter
between measurements on different devices, you should configure the
slaves to use the RTSI bus clock from the master NI 5102.
♦ NI 5102 (PXI)
The NI 5102 (PXI) uses the PXI Trigger <0..5> to carry RTSI Trigger
<0..5> and uses PXI Trigger 7 to carry the RTSI clock signal to all other
PXI slots in the system. RTSI Trigger 6 is reserved for use with PXI Star
Trigger, which the NI 5102 can receive but may not drive.
RTSI Trigger 0
TRIG
CH1
CH0
RTSI Trigger 1
RTSI Trigger 6
RTSI In 0
Analog Trigger
Circuit Output
Start Trigger
Stop Trigger
Scan Clock
Analog Trigger
Circuit Output
Digital
Mux
Analog
Trigger
Circuit
Analog Trigger Circuit Output
PFI1, PFI2
Internal Scan
RTSI In <0..6>
Scan Clock
Scan
Clock
Selection
PFI1, PFI2
RTSI In <0..6>
Analog Trigger Circuit Output
Stop Trigger
Stop
Trigger
Selection
PFI1, PFI2
Software
RTSI In <0..6>
Analog Trigger Circuit Output
Start Trigger
Start
Trigger
Selection
7
2
2
2
7
7
RTSI In 1
Start Trigger
Stop Trigger
Scan Clock
Analog Trigger
Circuit Output
Digital
Mux