Master/Slave Operation
You can use two or more NI 5102 digitizers in one system to increase the
number of channels for your application by synchronizing devices over the
RTSI bus or PFI lines.
Use the RTSI bus for synchronizing two or more NI 5102 (PCI, PXI, ISA)
devices. For the NI 5102 (PCMCIA, USB), you must use the PFI lines.
Restrictions
To ensure proper master/slave operation of your NI 5102, you must observe
the following restrictions:
You must use all channels for acquisition. For example, if you want to
use three channels at a time, you cannot use two channels on the master
and one channel on the slave, you must use four channels for data
acquisition and discard data on the fourth channel.
The desired pretrigger number of scans and total number of scans must
be a multiple of four. This is a hardware limitation.
There is a maximum of one sample clock timing jitter between master
and slave devices.
Connecting Devices
♦ NI 5102 (PCI, PXI, ISA)
You can synchronize NI 5102 (PCI, PXI, ISA) devices over the RTSI bus.
You can configure a system where a NI 5102 (PCI or ISA) can be the
master device controlling a mix of NI 5102 (PCI, ISA) slave devices. The
NI 5102 (PXI) can control only other NI 5102 (PXI) devices. The NI 5102
(PXI) has the RTSI connectivity on the bus connector; the optional
RTSI bus cable is not needed. However, you need a RTSI bus cable to
synchronize two or more NI 5102 (PCI or ISA) digitizers over the RTSI bus
as follows:
1.
If you are using multiple NI 5102 (PXI) digitizers, skip this step. If you
are using the NI 5102 (PCI or ISA), connect the master device with the
slave device over the RTSI connector. The cable and connector are
keyed so there is only one way to insert the cable in the connector.
2.
Ensure that no other device in the system is configured to output its
internal timebase on the RTSI bus clock line. The safest approach is
to restart your system, if possible.