Troubleshooting the Self-Test Verification
If the Self-Test Verification did not verify the chassis configuration, 
complete the steps in this section to troubleshoot the SCXI configuration.
Troubleshooting in NI-DAQmx
If you get a Verify SCXI Chassis message box showing the SCXI 
chassis model number, Chassis ID: x, and one or more messages 
stating Slot Number: x Configuration has module: SCXI-XXXX 
or 1125, hardware in chassis is: Empty, take the following 
troubleshooting actions:
Make sure the SCXI chassis is powered on.
Make sure all SCXI modules are properly installed in the chassis. 
Refer to the SCXI Quick Start Guide for proper installation 
instructions.
Make sure the cable between the SCXI chassis and E/M Series 
DAQ device is properly connected.
Inspect the cable connectors for bent pins.
Make sure you are using the correct NI cable assembly.
Test the E/M Series DAQ device to verify it is working properly. 
Refer to the E/M Series DAQ device help file for more 
information.
If you get a Verify SCXI Chassis message box showing the SCXI 
chassis model number, Chassis ID: x, and the message Slot 
Number: x
 Configuration has module: SCXI-XXXX or 1125, 
hardware in chassis is: SCXI-YYYY
, 1125, or Empty, complete the 
following troubleshooting steps to correct the error.
1.
Expand the list of NI-DAQmx devices by clicking the next to 
NI-DAQmx Devices.
2.
Right-click the SCXI chassis and click Properties to load the 
chassis configurator.
3.
Under the Modules tab, ensure that the cabled module is listed in 
the correct slot.
4.
If the cabled module is not listed in the correct slot, complete the 
following troubleshooting steps:
a.
If the cabled module is not listed in the correct slot and the 
slot is empty, click the drop-down listbox next to the correct 
slot and select the cabled module. Configure the cabled