Chapter 5    Using the SCXI-1125
© National Instruments Corporation
5-23
Perform Offset Null Compensation Using Traditional NI-DAQ (Legacy) 
C API
After configuring the system settings and module properties, you 
can perform an offset null compensation programmatically using 
SCXI_Calibrate
SCXI_Calibrate
 takes measurements and adjusts the 
coarse and fine offset null potentiometers to minimize or eliminate any 
electrical offset for a channel. Repeat this process for each channel by 
calling the 
SCXI_Calibrate
 function in a loop. Use the resulting 
imbalance in your application as a software correction factor by 
determining the residual voltage from the imbalance, and subtracting this 
residual offset from each future measurement. For more information 
regarding the operation of 
SCXI_Calibrate
, refer to the Traditional 
NI-DAQ (Legacy) Function Reference Help installed by default in Start» 
Programs»National Instruments»NI-DAQ
.
Perform Acquisition Using Traditional NI-DAQ (Legacy) C API
There are several NI-DAQ functions you can use to take measurements. 
Usually in SCXI the preference is to take multiple samples from multiple 
channels using the 
SCAN_Op
 function. 
SCAN_Op
 performs a synchronous, 
multiple-channel scanned data acquisition operation. 
SCAN_Op
 does not 
return until Traditional NI-DAQ (Legacy) acquires all the data or an 
acquisition error occurs (MIO, AI, and DSA devices only). For this reason, 
it is sometimes useful to use 
SCAN_Op
 in conjunction with the function 
Timeout_Config
, which establishes a timeout limit synchronous 
functions to ensure that these functions eventually return control to your 
application. After acquiring data using 
SCAN_Op
, the resultant data is not 
organized by channel, so you should demultiplex the data using 
SCAN_Demux
SCAN_Demux
 rearranges, or demultiplexes, data acquired by 
SCAN_Op
 into row-major order, meaning each row of the array holding 
Table 5-8.  NI-DAQ Functions Used to Configure SCXI-1125 
Channel 
Setting
NI-DAQ Function to Use
Significant 
Parameters
Possible Parameters 
Values
Gain
SCXI_Set_Gain
f64 gain
 
(gain setting)
1, 2, 5, 10, 20, 50, 100, 
200, 250, 500, 1000, 
2000
Bandwidth
SCXI_Configure_Filter
f64 freq
 
(filter cutoff 
frequency if 
filterMode = 1)
4.0, 10,000.0 Hz