Chapter 1
Installing and Configuring the SCXI-119X
1-6
ni.com
Note
When installing the SCXI-119module in an SCXI chassis, fill slot 1 first, then fill
slot 2 with the next SCXI-119X.
5.
Insert any other SCXI modules into the remaining slots in the same
manner as described in step 4.
6.
Secure all the SCXI-119modules to the SCXI chassis by tightening
both module thumbscrews.
Connecting the SCXI-119to a DAQ Device or Computer-Based
Instrument
You need the following items:
SCXI chassis with the SCXI modules installed (Refer to the
section for more
information.)
SCXI cable assembly, which consists of a cable adapter and a cable
An installed National Instruments DAQ device or computer-based
instrument, such as the NI 4060, that is able to control the SCXI
chassis
1/4 in. flathead screwdriver
Consult your SCXI chassis documentation, other SCXI module
documentation, and DAQ device or computer-based instrument
documentation for additional instructions and warnings. Your SCXI-119X
module and any other SCXI modules should already be installed in the
chassis according to their installation instructions.
You must select one of the modules in the chassis as the cabled
module—the module to which the SCXI cable assembly connects. If the
chassis contains at least one SCXI-1140 module, select an SCXI-1140 as
the cabled module; otherwise, select one of the remaining modules.
Note
The SCXI-119modules operate only in multiplexed mode.