SCC-TC Series Thermocouple Input Modules User Guide
8
ni.com
Cold-Junction Sensor
The cold-junction sensor voltage output varies from 1.91 to 0.58 V over a 
0 to 55 
°C temperature range.
Note
NI-DAQmx includes thermistor conversion utilities that implement the equations 
listed below. Refer to your software documentation for more information on these utilities.
You can use the following formulas to convert the cold-junction sensor 
voltage to cold-junction temperature:
where T
K
 is the temperature in degrees kelvin.
where
a = 1.295361 
× 10
–3
b = 2.343159 
× 10
–4
c = 1.018703 
× 10
–7
R
T
 is the resistance of the thermistor in ohms.
where V
TEMPOUT
 is the output voltage of the cold-junction sensor.
where T(°F) and T(°C) are the temperature readings in degrees Fahrenheit 
and Celsius, respectively.
Note
V
TEMPOUT
 varies from 1.91 V (at 0 °C) to 0.58 V (at 55 °C). For best resolution, use 
the maximum gain for this range on the analog input channel.
For the cold-junction sensor measurement accuracy, refer to the 
 section.
For more information about how to configure the SCC-TC0X using 
NI-DAQmx, refer to the SCC Quick Start Guide.
T
°C
(
)
T
K
273.15
=
T
K
1
a b
R
ln
T
(
c
R
ln
T
(
)
3
+
+
[
]
--------------------------------------------------------------
=
R
T
5,000
V
TEMPOUT
2.5 V
TEMPOUT
--------------------------------------
=
T
°F
( )
T
°C
(
) ]9
[
5
------------------------
32
+
=