Chapter 4
Theory of Operation
© National Instruments Corporation
4-11
Figure 4-8.  Unit Step Response of the SCXI-1141/1142/1143 Module
Setting the Cutoff Frequency
The cutoff frequencies of the filters in the SCXI-1141/1142/1143 module 
are set internally by dividing a base frequency of 100 kHz by an integer. 
You can determine the allowable cutoff frequencies for the 
SCXI-1141/1142/1143 module as follows:
where n is an integer 
≥ 4 and f
c
 
≥ 10 Hz. In other words, f
c
 = {25, 20, 16.7, 
14.3, 12.5, ..., 0.01} kHz.
If you are using NI software, the software automatically chooses a divisor, 
n, that best matches the cutoff frequency you specify and returns the actual 
cutoff frequency chosen.
The correct cutoff frequency depends on the application. If phase 
nonlinearity, ringing, passband ripple, or aliasing is a concern in the 
application, you may need to set the cutoff frequency several times higher 
0
2
4
6
8
10
0
0.2
0.4
0.6
0.8
1
1.2
1.4
Time (Seconds)
Bessel (1142)
Butterworth (1143)
Elliptic (1141)
Output V
oltage 
(V)
f
c
100
n
---------
 kHz
=