National Instruments cFP-RTD-122 用户手册

下载
页码 15
 
323348B-01
April 2003
© 2002–2003 National Instruments Corp. All rights reserved.
FieldPoint™, National Instruments™, NI™, and ni.com™ are trademarks of National Instruments Corporation. 
Product and company names mentioned herein are trademarks or trade names of their respective companies. 
For patents covering National Instruments products, refer to the appropriate location: Help»Patents in your software, 
the patents.txt file on your CD, or ni.com/patents.
FieldPoint Operating Instructions
FP-RTD-122 and 
cFP-RTD-122
Eight-Channel Three-Wire RTD and 
Resistance Input Modules
These operating instructions describe how to install and use the 
National Instruments FP-RTD-122 and cFP-RTD-122 three-wire 
RTD and resistance input modules (referred to inclusively as the 
[c]FP-RTD-122). For details on configuring and accessing the 
[c]FP-RTD-122 over a network, refer to the user manual for the 
FieldPoint network module you are using.
Features
The [c]FP-RTD-122 is a FieldPoint RTD and resistance input 
module with the following features:
Inputs for 100 and 1,000 
Ω platinum RTDs (resistance 
temperature detectors)
Built-in linearization for six TCR (temperature coefficient of 
resistance, or alpha) values of RTDs
Direct resistance measurements in 400 and 4,000
Ω ranges
True three-wire compensation
16-bit resolution
Filtering against 50 and 60 Hz noise
Hot swappable
2,300 V
rms
 transient overvoltage protection
–40 to 70 °C operation