Appendix B
Common Questions
© National Instruments Corporation
B-3
Analog Input and Output
Why is there a minimum sampling rate on the NI 6115?
The NI 6115 makes use of a pipelined ADC in order to achieve high 
sampling rates. Sampling at rates below 20 kS/s can result in improper 
digitization, which appear as noise in the acquired data.
How do I enable the programmable antialiasing filter on the 
NI 6115/6120?
In LabVIEW, select Data Acquisition»Analog Input»Advanced Analog 
Input»AI Parameter.vi 
from the function palette to set the filter values of 
50 kHz and 500 kHz for the NI 6115, or to enable the 100 kHz filter for the 
NI 6120, on a per channel basis. To disable the filter, set the filter value to 
0
.
Figure B-1.  Setting Filter Values in LabVIEW
In NI-DAQ, use the 
AI_Change_Parameter
 function to set the filter 
value. Set paramID to 
ND_Digital_Filter
. Set ParamValue to 
ND_High
 for a filter value of 500 kHz on the NI 6115 or 100 kHz on the 
NI 6120. Use 
ND_Low
 for a filter value of 50 kHz on the NI 6115. Use 
ND_None
 to disable the filter. The filter is disabled by default.
I have connected a differential input signal, but my readings are 
random and drift rapidly. What is wrong?
Check your ground-reference connections. The signal may be referenced 
to a level that is considered floating with reference to the device ground 
reference. Even if you are in differential mode, the signal must still be 
referenced to the same ground level as the device reference. There are 
various methods of achieving this reference while maintaining a high 
common-mode rejection ratio (CMRR). These methods are outlined in 
Chapter 4, 
.