Chapter 4
Connecting Signals
© National Instruments Corporation
4-25
 
Figure 4-19.  STARTSCAN Output Signal Timing
The CONVERT* pulses are masked off until the device generates 
STARTSCAN. If you are using internally generated conversions, the first 
CONVERT* appears when the onboard sample interval counter reaches 
zero. If you select an external CONVERT*, the first external pulse after 
STARTSCAN generates a conversion. STARTSCAN pulses should be 
separated by at least one scan period.
A counter on the NI 6115/6120 internally generates STARTSCAN unless 
you select some external source. This counter is started by the TRIG1 
signal and is stopped by either software or the sample counter.
Scans generated by either an internal or external STARTSCAN signal are 
inhibited unless they occur within a DAQ sequence. Scans occurring within 
a DAQ sequence may be gated by either the hardware signal AIGATE or 
the software command register gate.
t
w
t
= 25 – 50ns
a.  Start of Scan
b.  Scan in Progress, Two Conversions per Scan
Start Pulse
CONVERT*
STARTSCAN
t
off
 = 10 ns minimum
t
off