Chapter 1
Introduction and Quick Start
1-8
ni.com
Installing and Loading the NI-VXI/NI-VISA Software for Linux
For information on installing, uninstalling, or using the NI-VXI/NI-VISA 
software, refer to Chapter 5, 
.
VME Users
When used with a VXI-MXI-2, Resman identifies and configures the 
VXI devices, including the VXI-MXI-2. When used with a VME-MXI-2, 
Resman configures the VME-MXI-2 to allow the PCI-MXI-2 to access 
devices in the VME chassis. Resman does not configure VME devices. The 
VME specification does not specify the initialization and configuration 
procedures that the VXI specification requires.
It is recommended that you enter the information about your VME devices 
into the 
vxitedit
 utility. Resman can then properly configure the various 
device-specific VME address spaces and VME interrupt lines. For more 
information on configuring non-VXI devices in your VXI system, refer to 
Chapter 3, VXI Text Resource Editor, of the NI-VXI Text Utilities Reference 
Manual
. For more details about installing the NI-VXI software, refer to 
Chapter 5, 
, in this manual.
Device Interaction
After Resman has detected and configured all VXI/VME devices, you can 
view specific information on each device in your system by using the 
vxiedit
 utility or its text-mode counterpart, 
vxitedit
. These utilities 
include a Resource Manager Display, which contains a description for each 
device, including each VXI device’s logical address.
You can interact with your VXI/VME devices by using the 
vic
 and 
victext
 utilities for NI-VXI. These utilities let you interactively control 
your VXI/VME devices without using a conventional programming 
language, LabVIEW, or LabWindows/CVI.
Try the following in 
vic
:
1.
Click the Bus Access tab at the top of the window and select InReg as 
the Operation along the left side of the window.
2.
Select the VXI-MXI-2 or VME-MXI-2 as the Device Name under 
Input Parameters and the Id/Logical Address register under Input 
Parameters
.
3.
Click Go!. If the CMPL light along the right side of the window is 
green, and the output value ends in an FF6, you have successfully read 
the manufacturer ID for National Instruments.