Chapter 6
NI-VXI Configuration Utility
© National Instruments Corporation
6-29
Transfer Limit
You can use this feature to control how many data transfers the 
VXI/VME-MXI-2 will perform on the MXIbus before releasing it to 
another master device that is requesting use of the bus. The default setting 
holds the MXIbus for an unlimited period of time.
The other options you can choose from are 16, 64, and 256 transfers. If you 
do not want the VXI/VME-MXI-2 to hold the MXIbus for an unlimited 
period of time, you can use this control to select one of these values.
Parity Checking
By default, MXIbus parity checking is enabled and should not be disabled 
under normal circumstances. MXIbus parity is always generated regardless 
if checking is enabled or disabled.
MXI Fair Requester
You can use the Operate as Fair Requester checkbox control to configure 
the VXI/VME-MXI-2 as either a fair or unfair requester on the MXIbus. In 
its default setting (disabled), the VXI/VME-MXI-2 will request the bus at 
any time. If you enable this option, the VXI/VME-MXI-2 will request the 
MXIbus only when there are no requests pending from other MXIbus 
masters. This prevents other MXIbus masters from being starved of 
bandwidth. 
MXI CLK10 Signal
The VXI-MXI-2 can either receive or drive the MXIbus CLK10 signal. 
In its default setting, the VXI-MXI-2 uses the switch setting of S7 to 
determine the signal direction. 
♦ VME users—This option is not applicable to the VME-MXI-2.
You can use the Drive or Receive options to override the setting of S7 and 
control the direction of the MXIbus CLK10 signal. When receiving the 
MXIbus CLK10 signal, configure the W3 jumper setting to use the MXIbus 
as the source for generating the VXIbus CLK10 (applicable only if the 
VXI-MXI-2 is a Slot 0 device). When driving the MXIbus CLK10, the 
VXIbus CLK10 is used as the source. In this case, change the jumper 
setting so that it does not use the MXIbus CLK10 as the source for the 
VXIbus CLK10.
Caution
Do not configure more than one MXIbus device to drive MXI CLK10. Setting up 
a second device to drive MXI CLK10 could damage the device.