Chapter 5
Calibration
5-2
© National Instruments Corporation
If you use the PCI-1200 with NI-DAQ, LabVIEW, LabWindows/CVI, 
or other application software, the factory calibration constants are 
automatically loaded into the calibration DAC the first time a function 
pertaining to the PCI-1200 is called, and again each time you change your 
configuration (which includes gain). You can, instead, choose to load the 
calibration DACs with calibration constants from the user areas in the 
EEPROM or you can recalibrate the PCI-1200 and load these constants 
directly into the calibration DACs. Calibration software is included with 
the PCI-1200 as part of your NI-DAQ software. 
Calibration at Higher Gains
The PCI-1200 has a maximum gain error of 0.8%. This means that if the 
board is calibrated at a gain of 1 and if the gain is switched to 100, a 
maximum error of 32 LSB may result in the reading. Therefore, when you 
are recalibrating the PCI-1200, you should perform gain calibration at all 
other gains (2, 5, 10, 20, 50, and 100), and store the corresponding values 
in the user-gain calibration data area of the EEPROM, thus ensuring a 
maximum error of 0.02% at all gains. The PCI-1200 is factory-calibrated at 
all gains, and NI-DAQ automatically loads the correct values into the 
calibration DACs whenever you switch gains.
Calibration Equipment Requirements
The equipment you use to calibrate the PCI-1200 should have a ±0.001% 
rated accuracy, which is 10 times as accurate as the PCI-1200. However, 
calibration equipment with only four times the accuracy as the PCI-1200 
and a ±0.003% rated accuracy is acceptable. The inaccuracy of the 
calibration equipment results only in gain error; offset error is unaffected.
Calibrate the PCI-1200 to a measurement accuracy of ±0.5 LSBs, which is 
within ±0.012% of its input range.
For analog input calibration, use a precision DC voltage source, such as a 
calibrator, with the following specifications.
Voltage
0 to 10 V
Accuracy
±0.001% standard
±0.003% acceptable