SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
HI933700
Range 0 to 2000 µS/cm (HI 933700/1, HI933700/2);
0 to 4000 µS/cm (HI 933700/3, HI933700/4).
Accuracy (@ 20°C/68°F)
±5% f.s.
Setpoint
Adjustable from 0 to 100% f.s.
Setpoint Hysteresis
±10% f.s.
Alarm
Red LED lit if EC value exceeds SP (/1, /3)
Red LED lit if EC value is lower than SP (/2, /4)
Calibration
Manual with one trimmer
Probe
HI 7682 EC probe (included)
Power supply
HANNA Power Adapter (included)
Dimensions 79 x 49 x 95 mm (3.11 x 1.93 x 3.74”)
Weight
150 g (5.3 oz.)
Installation cat.
II
Output relay
1A max resistive load;
0.75 A max. inductive load; 250 VAC
INITIAL PREPARATION
INITIAL PREPARATION
INITIAL PREPARATION
INITIAL PREPARATION
INITIAL PREPARATION
• Connect the EC probe to the RCA
connector marked "PROBE".
• Connect the power adapter cable to
the power socket. The green front
panel LED should light.
• Connect the relay output contacts using the terminal NC
or NO according to your needs.
• Immerse the EC probe in the solution,
making sure that metal pins are com-
pletely submerged.
Note: the probe tip should be completely
submerged in the solution. More-
over, in order not to affect the accuracy of
measurements, it is important that the probe body
does not touch nor stand close to the side walls of the
beaker. The tip can lay on the bottom of the beaker.
• Wait for a couple of minutes for thermal equilibrium.
• Tap the probe gently on the bottom, then shake it while
rotating to make sure no air bubbles have remained
trapped.
• Adjust setpoint at 70%.
• Adjust the calibration trimmer
with a small screwdriver until the
output relay switches.
• Rotate the calibration trimmer in
the opposite direction until the
relay switches again.
• Adjust the calibration trimmer in
the middle of the two relay
switching points.
• The calibration is now complete
and the instrument is ready for use.
The instrument should be recalibrated at least once a month
and after performing probe cleaning procedure.
For the greatest accuracy, frequent calibration of the instru-
ment is recommended.
• Turn the meter on and connect the probe.
• Pour a small quantity of
HI 7031 EC calibration
solution (for HI 933700/
1 and /2) or HI 7032
calibration solution (for
HI933700/3 and /4) in
a beaker. If possible, use
plastic beakers to minimize
any EMC interference.
CALIBRATION
CALIBRATION
CALIBRATION
CALIBRATION
CALIBRATION
ACCESSORIES
ACCESSORIES
ACCESSORIES
ACCESSORIES
ACCESSORIES
HI 7682
EC probe and 1 m (3.3’) cable
HI 70031P 1413 µS/cm (EC) calibration solution, 20 mL
sachet (25 pcs)
HI 70032P 1382 ppm/cm (EC) calibration solution, 20 mL
sachet (25 pcs)
HI 7031L 1413 µS/cm (EC) calibration sol., 460 mL
bottle
HI 7032L 1382 ppm calibration sol., 460 mL bottle
HI 7061L Electrode cleaning solution, 460 mL bottle
HI 710005 12 VDC power adapter, US plug
HI 710006 12 VDC power adapter, European plug
HI 710012 12 VDC power adapter, Australian plug
HI 710013 12 VDC power adapter, Southern Africa plug
HI 710014 12 VDC power adapter, UK plug
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
• Select your own set point via the front panel setting
knob.
• Immerse the tip of the EC probe
in the sample.
Note: In order not to affect the
accuracy of measurements, it
is important that the probe
body does not touch nor stand
close to the side walls of the
vessel. The tip may lay on the
bottom of the beaker.
• When the reading of the controller be-
comes higher or lower than the setpoint
(depending on the model), the red, front panel LED
illuminates and the output relay is activated.
PROBE MAINTENANCE
To minimize clogging and extend the life of the probe, it is
recommended that it be cleaned often or at least once a
month.
• Immerse the tip of the electrode in HI 7061 Cleaning
Solution for one hour.
• If a more thorough cleaning is required, rub the metal
pins with very fine sandpaper.
• After cleaning or when not in use, rinse the probe with tap
water and recalibrate the instrument.
CE DECLARATION
CE DECLARATION
CE DECLARATION
CE DECLARATION
CE DECLARATION
OF CONFORMITY
OF CONFORMITY
OF CONFORMITY
OF CONFORMITY
OF CONFORMITY
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environ-
ment in which they are used.
Operation of these instruments in residential areas could cause unacceptable interfer-
ences to radio and TV equipment.
The metal band at the end of the probe is sensitive to electrostatic discharges. Avoid touching
this metal band at all times.
During operation, ESD wrist straps should be worn to avoid possible damage to the probe by
electrostatic discharges.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s
EMC performance.
To avoid electrical shock, do not use these instruments when voltages at the measurement
surface exceed 24 VAC or 60 VDC.
Use plastic beakers to minimize any EMC interferences.
To avoid damages or burns, do not perform any measurement in microwave ovens.
The calibration trimmer is sensitive to electrostatic discharges. It is recommended to use
antistatic screwdriver.
ISTR933700  03/98