rototherm c105 temperature recorder 用户手册

下载
页码 16
3
Specification
General:
Chart diameter
255mm
Writing width
105mm
Intrinsic error
± 1% span maximum (±2% R.H. for hygroscopic
membrane system)
Operating temperature limits
–10
o
to +50
o
C
Except for
Operating humidity limits
0 to 80% R.H.
humidity recorders
Zero error due to ambient temp. variations
±0.05% span/
o
C typical
Chart speeds
1 rev every 12 hours, 24 hours or 7 days
Chart drive
Synchronous electric motor or mechanical clock
Power supply voltage and frequency
200/250V or 100/120V, 50 or 60Hz
Mounting
Wall or panel
Temperature measuring instruments:
Measuring systems
Fluid expansion
Maximum length capillary
30m
Minimum span
40
o
C
Maximum span
500
o
C
Span limits
– 30
o
C to +600
o
C
Pressure measuring instruments:
Measuring system
Capsule stack, Bourdon tube or coiled tube
Minimum span
0 to 1 bar vacuum or pressure.
Maximum span
1200 bar
Span limits
–1 bar to 1200 bar
Connection size
Ranges up to 40 bar: 
inch B.S.P.) with nut and tail
Ranges over 40 bar: 
2
inch B.S.P.) piece for 8mm tube
Humidity measuring instruments:
Measuring systems
Sensitive hygroscopic membrane and fluid filled system
for temperature measurement, or wet and dry bulb fluid
expansion systems.
Spans
20 to 100% R.H. (hygroscopic membrane);
0 to 40
o
C (fluid filled systems)
Operating temperature limits
0 to 40
o
C
Operating humidity limits
30 to 85% R.H.
Maximum length of capillary
(wet and dry bulb system)
30m (each bulb)
Electric alarm or control:
Instrument contacts
20mA inductive or 30mA non-inductive load max.
Contact action
Make above or below set point as required
Relays
Delayed action double pole change-over, 5A, 240V
50Hz non-inductive load, as standard
(Refer to wiring Diagram, p.14).
Relay action
Energised above or below set point as required.
Overall dimensions
350mm wide; 137mm deep; 350mm high
(Temp & R.H. 582mm high)
Panel cut-out
318mm diameter
Maximum panel thickness
25mm
Weight
10kg
66
77
88