Instruction Manual 
 
Bromide Electrode 
 
18
Low Slope or          
standards contaminated     
 
prepare fresh standards  
No Slope                   or incorrectly made     
                      
                         
ISA not used              
 
use recommended ISA 
 
                         
standard used as ISA    
 
use ISA 
 
                         
electrode exposed        
 
check section entitled 
                         
to complexing agents                  
Complexation  
 
                         
air bubble on membrane          
 
remove bubble by 
                         
 
 
 
 
 
re-dipping probe 
 
 
"Incorrect Answer"     incorrect scaling of     
 
plot millivolts on the linear 
(but calibration   
semi-log paper      
 
 
axis. On the log axis, be                                
curve 
is 
good) 
      sure 
concentration 
numbers 
                                          
 
 
 
 
within each decade are 
                                      
 
 
 
 
increasing with increasing 
                                                           
 
 
concentration. 
 
                         
incorrect sign          
 
 
be sure to note sign of  
 
        millivolt 
number 
correctly 
 
                         
incorrect standards  
 
 
prepare fresh standards 
 
                         
wrong units used          
 
apply correct correction factor: 
  
 
 
 
 
 
 
 
10
-3
M = 79.9 ppm as Br
-1
 
 
                         
complexing agents       
 
check section entitled  
                         
in sample                                 
Complexation