Lumiscope 1145 用户手册

下载
页码 20
14
 Problem  Symptom  Check This 
Remedy
Error 
message
Err shows 
on the 
display.
A measure-
ment error 
occurred
Relax for a 
moment and then 
measure again
Err 1, 2, or 
3 shows 
on the 
display
The wrist cuff 
is not secure
Refasten the cuff 
and then measure 
again
Err 4 
shows on 
the display
The monitor 
detected 
motion while 
measuring
Movement 
can affect the 
measurement.
Relax for a 
moment and then 
measure again
Err 5 
shows on 
the display
Pressure is 
over 37.3 kPa 
(280 mmHg)
Relax for a 
moment and then 
measure again
Err 6 
shows on 
the display
Deflation 
period was too 
long
Movement 
can affect the 
measurement.
Relax for a 
moment and then 
measure again
 Problem  Symptom  Check This 
Remedy
Error 
message
Err 7 or 8 
shows on 
the display
A calibration 
error occurred
Retake the 
measurement. If the 
problem persists, 
contact the retailer 
or our customer 
service department 
for further 
assistance. Refer 
to the warranty for 
contact information 
and return 
instructions
Settings 
are wrong
Date and 
time are 
incorrect
The clock was 
not set or reset 
after installing 
new batteries
Reset the clock
Measure-
ment unit 
(kPa or 
mmHg) are 
incorrect
The 
measurement 
unit was not set 
or reset after 
installing new 
batteries
Reset the 