VW40-152-009
 
VIVIX-S 1012N User Manual 
 
 
Be sure to select the Defect Map data file provided with a detector. If the file has a wrong 
serial number or you select a wrong file, an error message will be displayed. 
 
 
 
Vieworks provides the Defect Map data stored in the detector. However, you should 
download the data as the calibration cannot be performed in the detector automatically. 
 
This Defect Map data has been generated through stringent test in the production stages 
of detector. However, new defect may be newly formed while using the detector. In this 
case, make sure to calibrate defects in person and generate a new defect map. 
6.4.3  Loading Gain Calibration Data 
 
 
1  Select Gain Calibration – Normal from the Step list. 
2  Click Download Gain button to assign the path of local HDD, and download the Gain calibration data. 
 
 
 
The calibration menu status is displayed as Done when the Gain calibration data is 
completed to be loaded. 
 
 
 
Be sure to select the Gain calibration data file provided with a detector. If the file has a 
wrong serial number or you select a wrong file, an error message will be displayed. 
 
 
 
Vieworks provides the Gain calibration data stored in the detector. However, you should 
download the data as the calibration cannot be performed in the detector automatically. 
 
This data is provided in case the Gain calibration cannot be performed. You should do 
the Gain calibration in person as the detector condition can be different by the operating 
method or use environment.