Vieworks Co. Ltd. FXRD-1012NAW 用户手册

下载
页码 151
 
 
Rev.1.0
 
Page 33 of 151 
VW40-152-009
 
VIVIX-S 1012N User Manual 
2.3.5  Wireless Communication 
Item 
Specifications 
Wireless standard 
IEEE802.11n 
Frequency range 
2.412 ~ 2.472㎓  (13 Channels) 
5.18 ~ 5.24㎓  (4 Channels) 
5.745 ~ 5.805㎓  (4 Channels) 
Data transmission rate 
802.11n: Max. 300Mbps (MIMO 2X2) 
Modulation 
BPSK, QPSK, 16-QAM, 64-QAM 
Transmission power 
Max. 17dBm 
Security 
WPA2-PSK 
Antenna 
Dual Band Antennas (2EA, inbuilt) 
 
 
 
The specification of detector and SCU are same, but the location of antenna is different. 
2.3.6  Use Environment 
Item 
Operation 
Storage & Transportation 
Temperature 
+10 ~ +35℃ 
-15 ~ +55℃ 
Humidity 
30 ~ 85% (Non-condensing) 
10 ~ 90% (Non-condensing) 
Atmospheric pressure 
700 ~ 1060 hPa 
500 ~ 1060 hPa 
Shock 
1.6G 
20G 
Vibration 
0.7G 
0.7G 
 
 
 
The use environment of detector and SCU is same. 
 
Load Limit of Detector 
Uniform load 
Local load 
Over the whole surface 
Center diameter 40mm 
Max. 100 kg 
Max. 150 kg 
 
 
 
 
 
Do not let the paitent or object heavier than load limit be on the detector. Then, detector 
can be damaged. 
 
Do not let the patient lie or get on the detector. Internal devices such as a sensor can be 
seriously damaged even if his/her weight is within the load limit.