Starting Continuous Glucose Monitoring 
 
Choose a site on your abdomen (belly) to put the Sensor that is a flat area and where you don’t have any scarring or rough patches 
of skin and clean the site with an alcohol swab.  Remove the adhesive backing from the STS™ Sensor Pod and place the STS™ 
Sensor on the selected site. 
1.  Pull the safety lock out from the Applicator.  
2.  Place 2 fingers under the collar, with your thumb on the plunger, and push the plunger down completely. 
3.  At the same time that you are pushing the plunger down, pull the collar back completely until you hear a click. 
4.  Press the tabs on the sides of the STS™ Sensor Pod and pull out the STS™ Applicator.
 
 
 
 
 
 
 
 
4. 
S
TEP
 1- Inserting a New STS™ Sensor 
Push plunger 
Remove STS™ Applicator 
STS™ Sensor Pod remains adhered 
Press tabs 
Collar completely back 
Safety Lock 
1. 
Pull Collar 
2. 
3. 
4. 
 
STS™ Sensor &  STS™ Applicator 
STS™ Sensor & 
STS™ Transmitter 