Disable the PASS / FAIL Function 
The HHD PASS / FAIL function can be disabled by setting the threshold value to 
80 dB. In this mode, neither the PASS nor the FAIL indicator appears. 
 
Enable the PASS / FAIL Function 
To enable the HHD PASS / FAIL function, set the threshold value to a setting 
other than 80 dB. 
 
Test Signal Verification 
The combined PASSFAIL label is displayed along with the integrity reading if the 
HHD is unable to clearly receive the VMD test signal. The HHD performs a pretest 
signal level measurement prior to triggering the VMD and then compares that 
level with the measurement made during the VMD transmit interval. If the VMD 
signal level is no greater than the pretest level, then the combined PASSFAIL 
label is shown. The test signal verification function is not affected by disabling the 
Threshold function.