(3) Before Exposure
Be sure to check the equipment daily and confirm that it works
properly.
Sudden heating of the room in cold areas will cause
condensation to form on the instrument. In this case, wait until
condensation disappears before performing exposure.
Room temp. 70℉ / 21℃ is recommended.
If the instrument is used with condensation formed on it,
problems may occur in the quality of the instrument. When an air-
conditioner is going to be used, be sure to raise/lower the
temperature gradually so that a difference in temperature in the
room and in the instrument does not occur, to prevent forming of
condensation.
(4) During Exposure
Do not use the detector near devices generating a strong
magnetic field. Doing so may produce image noise or artifacts.
Do not apply excessive weight to the sensor unit. Otherwise, the
sensor may be damaged.