Rayence Co. Ltd. RY1417W 用户手册

下载
页码 83
51 
 
R-USM-020 
 
 
 
 
5.  Calibration 
 
(1)  General Principle  
Notation 
Calibration  can  be  done  by  image  acquisition  S/W.  The  gain-
offset correction (under calibration) will be done with one dark, at 
least one bright and object frame. 
 
 
 
 
 
 
 
 
 
 
Parameter 
Description 
Offset 
Dark image, acquired image without X-ray exposure 
Bright 
Acquired image with X-ray exposure 
Object 
Bright image with object, will be calibrated 
Gain 
Gain of imaging system, offset subtracted image 
Offset correction 
Offset subtract 
Gain correction 
Compensate gain variance of pixel 
                X-ray  
                detector  
                should be 
used at stable state 
within driving 
temperature range. 
Acquire the X-ray 
images after power 
on and 5 minutes 
warming up to 
obtain high quality 
images.