R-USM-020 
 
 
 
 
(3)  Inspection 
In order to ensure that the instrument is used safely and normally, 
please  be  sure  to  inspect  the  instrument  before  use.  If  any 
problem  is  found  during  the  inspection,  please  take  measures 
indicated  in  this  chapter.  If  problem  still  cannot  be  corrected, 
please  contact  Rayence  representative  or  distributor.  It  is 
recommended that a record of the inspection be kept by making 
copies  of  the  check  lists  in  this  section,  or  making  a  separate 
check list. 
Inspection chart 
Inspection 
User  Vendor 
Cycle 
Check  that  cables  are  not 
damaged  or  cover  of  cables  is 
not torn 
 
Daily 
Check  that  the  plugs  and  locks 
of connectors are not loose 
 
Daily 
Check  that  the  cover  or  parts 
are not damaged and not loose 
 
Daily 
Check the LED indicator 
 
Daily 
Check the bad pixel map 
 
Half year 
Check  the  performance  of  the 
instrument 
by 
performing 
exposures using a phantom or a 
resolution chart 
 
Yearly