Mother Transmitter
A Mother transmitter is placed on the wrist of a new mother and then linked to an Infant 
transmitter. When the Infant transmitter is in contact with the Mother transmitter, the 
Mother transmitter flashes Green LED to indicate that the Infant and Mother 
transmitters match. If there is a mismatch, meaning the Infant and Mother transmitters do not match, the 
Mother transmitter flashes the Red LED three times, and sounds a three beep alarm. A Mismatch Transmitter 
alarm is also generated in the Event List. If there is a linking problem, a Link Transmitter alarm is 
generated in the Event List. 
In cases where the Mother Transmitter is linked but no longer active (e.g. the patient has been 
discharged but the transmitter was never cleared) the transmitter will beep after 24 hours of inactivity. 
If the transmitter go undetected and allowed to beep for 15 minutes, it will clear and discharge itself 
from the system. Tapping the Mother Transmitter will temporality stop the beeping but the 
transmitter remains in a linked state until manually cleared (refer to the Mother Baby Match Mother 
Transmitter User Guide
). If not cleared, the transmitter will beep again after another 24 hours of 
inactivity.
N
OTE
 Asset transmitters can also be linked to Mother transmitters. When an asset 
transmitter is linked to the Mother transmitter, all linked functions are applicable.
CAUTION: The Mother Transmitter can be worn in the shower; complete 
submersion should be avoided. Prolonged submersion may result in equipment 
and/or system failure and voids any or all Product Warranty.