Transport and Storage Conditions
The Patient Assistant may not work correctly if it is exposed to heat, humidity,
or atmospheric-pressure levels outside the following ranges:
Short-term storage and transport conditions – If the Patient Assistant is
exposed to conditions in the following ranges for more than 24 hours, it may
stop working entirely:
• Temperatures of -25 to 5°C (-13 to 41°F) at a relative humidity up to 90%,
non-condensing
• Temperatures of 35 to 60°C (95 to 140°F)
Long-term storage conditions – The Patient Assistant is designed to be
exposed to conditions in the following ranges for up to 2 years while still
functioning:
• From 15 to 30°C (59 to 86°F) at a relative humidity up to 90%