Device Classifications 
 
SmartPill pH.p Capsule 
SmartPill Data Receiver 
SmartPill Docking Station 
Power: 
Internally Powered 
Equipment 
Internally Powered Equipment  Class I Equipment 
Electric Shock 
Protection: 
Type BF Applied Part 
Type BF Applied Part 
Ordinary Equipment 
Harmful Ingress 
Water: 
Watertight Equipment 
(IP57) 
Ordinary Equipment 
Ordinary Equipment 
 
Equipment not suitable for 
use in the presence of 
FLAMMABLE 
ANAESTHETIC 
MIXTURE WITH AIR or 
WITH OXYGEN OR 
NITROUS OXIDE. Not 
for use with oxygen or 
oxygen enriched 
atmospheres. 
Equipment not suitable for use 
in the presence of 
FLAMMABLE 
ANAESTHETIC MIXTURE 
WITH AIR or WITH 
OXYGEN OR NITROUS 
OXIDE. Not for use with 
oxygen or oxygen enriched 
atmospheres. 
Equipment not suitable for use 
in the presence of 
FLAMMABLE 
ANAESTHETIC MIXTURE 
WITH AIR or WITH 
OXYGEN OR NITROUS 
OXIDE. Not for use with 
oxygen or oxygen enriched 
atmospheres. 
Mode of 
Operation: 
Intermittent operation 
Continuous Operation 
Continuous Operation 
  
Connectivity Restrictions 
External equipment intended for connection to signal input, signal output or other 
connectors shall comply with IEC 60601 series for medical electrical equipment. In 
addition, all such combinations – systems – shall comply with the standard IEC 60601-
1-1, Safety requirements for medical electrical systems. Equipment not complying with 
IEC 60601-1 shall be kept outside the patient environment, as defined in the 
standard.5 
Any person who connects external equipment to signal input, signal output or other 
connectors has formed a system and is responsible for the system to comply with the 
requirements of IEC 60601-1-1. If in doubt, contact a qualified biomedical 
professional or the SmartPill Corporation. 
 
Index 
                                           
 
 
 
5 The normal distance is at least 1.5 m from the patient or the patient support.