Appendix: Electrical Safety 
was one additional esophageal retention, eight gastric retentions, four small bowel 
retentions and five colonic retentions.  These resolved either without intervention or 
with endoscopy/colonoscopy.   
Patient-Contacting Materials 
Patient-contacting materials include polyurethane, a polyurethane-polycarbonate blend, 
Teflon• coated with polyhema, an ISFET pH sensor, epoxy and UV-cured adhesive. 
All patient contacting materials have been tested for biocompatibility and have been 
found non-toxic, non-sensitizing, and non-irritating. 
The device does not contain natural rubber latex. 
Care, Cleaning and Maintenance 
Caution 
Do not reuse capsules. The capsule is a single-use, disposable item. 
Caution 
Do not immerse the data receiver in water or other liquids. Immersion damages internal electrical components 
resulting in data receiver inoperability. 
Caution 
Do not immerse the docking station in water or other liquids. Immersion damages internal electrical 
components and could result in electrical shock. 
Caution 
The data receiver, docking station, power supply, and capsule are not user serviceable. 
Data Receiver  
Clean and disinfect the outside surfaces of the data receiver after each patient use. 
1. Turn off data receiver by simultaneously depressing the backlight button and event 
button for 5 seconds. 
2. Wipe outside surfaces with a cloth dampened (not saturated) with a mild detergent 
and water. Suitable detergents include dishwashing detergent (e.g., Joy, Dawn®, 
Palmolive®) solutions, or laboratory glassware cleaners such as Alconox®. 
3. Wipe dry. 
4. Disinfect by wiping outside surfaces with a cloth dampened (not saturated) with 
disinfectant. A solution of 10% household bleach can be used as a disinfectant. 
5. Wipe dry. 
6. Wait at least 5 minutes after cleaning and disinfection before turning the data 
receiver on or placing it in the docking station. 
7. Conduct a preventive maintenance inspection. Check for cracks or damage.  Shake 
the data receiver and listen for detached batteries.  Do not use the data receiver if 
the case is cracked or damaged, or a rattling can be heard.