Xmark Corporation PT2 用户手册

下载
页码 39
EXI HALO User Manual
980-000002-000
 
EXI Wireless Systems Inc.
17
December 2000
Revision 6.0
 
Alarm type (tag removal or egress alarm)
 
Infant name
 
Tag number
 
Time
 
ID of staff member accepting the alarm
 
Staff member notes on alarm incident
2.5.  “Nurse Saver” and “Loitering” Features
 
If the system is installed with the “Nurse Saver” feature, Tags detected at egress points that
are secured will not set off nuisance alarms.
 
A Tag detected near a door that is closed will result in the appropriate floor button on the
computer screen turning yellow (pre-alarm), and the icon associated with that door also
turning yellow.
 
If a bypass keypad is installed at the door, it will flash a light and emit periodic “beeps” to
indicate the presence of the Tag.  This event is logged into the computer as “Tag detected
at location xxx”, but will not create an alarm condition.
 
If the door is opened while the Tag is still near the door an alarm will be initiated. This
feature can only operate if the door is equipped with a magnetic switch to sense whether it
is open or closed.
 
If the tag remains near the door for a period of one minute or more, an egress alarm will be
initiated and the button and the icon on the screen will both turn to red.  This is the “Loi-
tering” feature.
Pre-Alarm
(Yellow)
Alarm (Red)