MESSAGES AND INFORMATION 
APPENDI X A 
 
21 
The following table describes messages that are displayed on the WAM during patient hookup and transmission. 
 
Table of Messages 
 
Message 
Solution 
Solid Green LED 
No interaction required 
Flashing Green LED 
Replace AA battery 
Solid Yellow LED 
Lead fail message, check leads for proper connection 
Flashing Yellow LED 
Low battery and lead fail condition exist. Replace battery and check leads for proper 
connection 
LED Off 
Power is off press power button to turn on. If no audible beeping is heard, battery is 
completely depleted replace battery on power on WAM. If beeping is heard assure you 
are within 3 meters of the electrocardiograph. 
 
Following Message’s will be seen on the ELI 350 if applicable 
RA or N 
RA or N fail.  Check if the lead wire is off or the electrode needs to be replaced. 
RL or R 
RL or R fail.  Check if the lead wire is off or the electrode needs to be replaced. 
LA or F 
LA or F fail.  Check if the lead wire is off or the electrode needs to be replaced. 
LL or L 
LL or L fail.  Check if the lead wire is off or the electrode needs to be replaced. 
A combination of  
RA/
…/LL or N…F 
More than one limb lead fail or all leads fail.  Check the lead wires and electrodes. 
V1 or C1 
V/C1 fail.  Check if the lead wire is off or the electrode needs to be replaced. 
V2 or C2 
V/C2 fail.  Check if the lead wire is off or the electrode needs to be replaced. 
V3 or C3 
V/C3 fail.  Check if the lead wire is off or the electrode needs to be replaced. 
V4 or C4 
V/C4 fail.  Check if the lead wire is off or the electrode needs to be replaced. 
V5 or C5 
V/C5 fail.  Check if the lead wire is off or the electrode needs to be replaced. 
V6 or C6 
V/C6 fail.  Check if the lead wire is off or the electrode needs to be replaced. 
A combination of 
V1, V2, V3, V4, V5, V6 
or 
C1, C2, C3, C4, C5, C6 
More than one chest lead fails.  Check the lead wires and electrodes. 
Note:  Lead Wire messages will be seen on the ELI 350 display and not on the WAM.  The WAM will display a 
yellow LED in the event of a lead failure.